Measured value simulation, Simulation mass flow qm abs, Density simulation – Fluid Components International CMB User Manual

Page 87: Temperature measurement simulation

Advertising
background image

13.11.3 Measured value simulation


If the operator selected the setting “QM, D, T” described in Section 13.11.2 on page 86, the following
three possible settings will affect the output behavior during measured value simulation, where all meas-
ured values are simulated at the same time.

13.11.3.1 Simulation mass flow QM abs

In order to simulate mass flow, the operator can define a “measured value.” The flow rates will be simu-
lated in both directions. All outputs will perform based on the simulated measured value.

Set QM abs

±0900.0 kg/h


The simulation value is entered as described in Section 12.4.3.2 Input window/modify a value.

13.11.3.2 Density simulation

In order to simulate density/volume measurement, the operator can define a “density measured value.” If
volume measurement is assigned to an output, it will change depending on mass flow and density simu-
lation. All outputs will perform based on the simulated measured value.

Set density

0500.0 g/l


The simulation value is entered as described in Section 12.4.3.2 Input window/modify a value.

13.11.3.3 Temperature measurement simulation

In order to simulate a temperature, the operator can define a “measured value.” All outputs will perform
based on the simulated measured value.

Set temperature

+090 °C


The simulation value is entered as described in Section 12.4.3.2 Input window/modify a value.

CMB & CT OPERATING MANUAL