Direct simulation of outputs, Status output simulation, Pulse output simulation – Fluid Components International CMB User Manual

Page 88: Simulation current output i1, Simulation current output i2

Advertising
background image

13.11.4 Direct simulation of outputs


If the operator selected the setting “Direct simulation” described in Section 13.11.2 Direct simulation on
page 86, the following four possible settings will affect the output behavior during measured value simula-
tion, where all measured values are simulated at the same time.

13.11.4.1 Status output simulation

The Status output simulation function allows the operator to purposefully activate the status output.
Press

↵ to display the current state.

Status output

[off]


As mentioned in Section 12.4.3.1 Selection window/make a selection, the operator can toggle between
“on” and “off.”

13.11.4.2 Pulse output simulation

The Pulse output simulation function allows the operator to define a frequency to be assigned to the pulse
output. After selecting this function and pressing

↵, the following selection field will be displayed:

Set frequency

0210.0 Hz


This field shows the current frequency. As mentioned in Section 12.4.3.2 Input window/modify a value,
the definable frequency ranges from 6 Hz to 1100 Hz.

13.11.4.3 Simulation current output I1

This function allows the operator to define a current for current interface 1. Press

↵ to display the set

current.

Set I1

I1 = 10.50 mA


As mentioned in Section 12.4.3.2 Input window/modify a value, the current value can be changed.

13.11.4.4 Simulation current output I2

As described in Section 13.11.4.3, current output 2 can also be configurated.


Page 88 of 106

CMB & CT OPERATING MANUAL