13.10.1 Current output I1 0/4 to 20 mA

The Current output I1 0/4 to 20 mA function allows the operator to define the range in which the current
output is to be operated. Within the range from 0 to 21.6 mA (= 0 ... 110 %) HART communication

®

is not

possible. The range from 4 to 20.5 mA follows the NAMUR recommendation and covers the range from 0
to 104 % of the measuring range. The standard range from 4 to 21.6 mA allows for a control of the meas-
uring range of up to 110 %.

Press

↵ to display the current setting.

Current output I1

[4] – 21.6 mA


As mentioned in Section 12.4.3.1 Selection window/make a selection, the operator can choose between
the following settings:

Current output

0,00 mA

5,00 mA

10,00 mA

15,00 mA

20,00 mA

25,00 mA

0%

20%

40%

60%

80%

100%

120%

Measured value

Out

put

c

u

rre

nt

0 - 20 (21,6) mA

4 - 20,5 mA
NAMUR

4 - 21,6 mA
Standard

13.10.2 Current output I1 alarm

This function allows the operator to define the state taken on by the current output when a state of alarm
is detected. This information can be analyzed in the control system. Press

↵ to display the current setting:

I1 : alarm

[>22mA]



As mentioned in Section 12.4.3.1 Selection window/make a selection, the operator can choose between
the following settings:

¾

not used

no alarm function

¾

> 22 mA

current rise in the case of an alarm

¾

< 3.8 mA

current reduction in the case of an alarm



13.10.3 Current output I1 assignment