Designation

Terminal designation

Type of protection

Standard

EEx ia

EEx e

(Not Ex)

Signal outputs

11 and 12

x

x

Current 1, 0/4 to 20mA
with HART

®

41 and 42

x

13 and 14

x

x

Current 2, 0/4 to 20mA

43 and 44

x

16 and 17

x

x

Binary output 1
(passive pulse)

46 and 47

x

45 and 48

x

Binary output 1
(active pulse)

15 and 18

x

19 and 20

x

x

Binary output 2
(status or second pas-
sive pulse output for
custody transfer opera-
tions)

49 and 50

x

33 and 34

x

x

Option
Binary output 3
(status output during
custody transfer opera-
tions)

53 and 54

x

Profibus PA

option

39 (A) and 40 (B)

x

Control unit BE

Shield, -, +

x

x

Alternatives for cur-
rent output 2

21 and 22

x

x

Binary input

51 and 52

x

Modbus/Profibus DP
with RS 485-IS

35 (A) and 36 (B)

(not currently available)

x x

Profibus DP

37 (A) and 38 (B)

(not currently available)

x

¾

Since the signal outputs cannot be activated simultaneously owing to the limited number of ter-

minals available, one of the aforementioned options must be selected. Field bus devices
(Profibus PA) are not outfitted with an analog or impulse output.

¾

A maximum of 8 signal output terminals is available (in addition to the terminal for the control unit

and Profibus PA).

¾

“Increased safety” type of protection signal outputs are to be connected only to "Extra low volt-

age" degree of protection circuits with safe electrical isolation in accordance with DIN VDE 0100
part 410.

¾

Under no circumstances are “Increased safety” signal outputs to be combined with “In-

trinsic safety” signal outputs

.

¾

If interface output RS 485 (under development and not currently available), which is only avail-

able in an “Increased safety” model, is selected, all signal outputs must also have this type of pro-
tection.

¾

If “Control input” or “RS 485 interface” are selected, current output 2 is not supplied.

¾

If the sensor and transmitter are interconnected using a cable, the following cable is to be used:

SLI2Y (SP) CY 5 x 2 x 0.5 mm (for explosion-proof applications, grey for non-explosion
proof applications).