Current output i2 alarm, Current output i2 assignment – Fluid Components International CMB User Manual

Page 84

Advertising
background image

13.10.5 Current output I2 alarm


This function allows the operator to define the state taken on by the current output when a state of alarm
is detected. This information can be analyzed in the control system. Press

↵ to display the current setting.

I2 : alarm

[not used]



As mentioned in Section 12.4.3.1 Selection window/make a selection, the operator can choose between
the following settings:

¾

not used

no alarm function

¾

> 22 mA

current rise in the case of an alarm

¾

< 3.8 mA

current reduction in the case of an alarm




13.10.6 Current output I2 assignment

This function allows the operator to define the measured value to be output as an analog signal via cur-
rent output I2. Press

↵ to display the current setting.


I2 assigned to
[Temperature]


As mentioned in Section 12.4.3.1 Selection window/make a selection, the operator can choose between
the following settings:

¾

Mass

flow

¾

Volume

flow

¾

Density

¾

Temperature

¾

not available (in this case the vendor setting must not be changed)


Page 84 of 106

CMB & CT OPERATING MANUAL

Advertising