R Series Service Manual

9650-0903-01 Rev. J

5-5

Accessories

R Series accessories include the following:

All legacy M Series accessories (paddles, padz, internal handles, SpO

2

, 3/5Lead ECG, NIBP cuffs, etc.)

EtCO

2

Capnostat 5 sensor.

R Series unicables and R Series electrodes. The fully featured models support defib, pace, CPR, date code, condition sensor, defib self-test while pre-
connected, and monitoring while pacing.

Sync in/Sync out cable (Not sold by ZOLL - third party makes this cable).

USB cable (Not sold by ZOLL - compatible with standard USB cables).

Printer paper.

Power Management Support Functions

Charger Functions

When battery is fully charged, the charge current is terminated by setting the charger current PWM to less than 15%. The charger resumes supplying
battery current when the SOC reaches 95% +/-1%.

R Series powers on when a battery is dropped-in if the loaded battery Voltage is above the hardware threshold (8.7V). Upon battery drop-in, charger
current-limit is established on the basis of Battery conditions and DC bus Voltage. If the battery charge was recently terminated due to over-current, the
charger FET is opened, and the replace battery is indicated.

Shut-down Functions

The software initiates the shutdown sequence (“replace battery”) when the Kelvin Battery A/D Voltage is 9.3V or at 600mA-hrs of remaining capacity,
whichever occurs first. (Remaining capacity is read from the battery as a 16-bit unsigned quantity directly in mA-hrs).

The hardware will not shutdown if the input Voltage (at the switched-power node) is above 8.7V.

The software initiates “low battery” at an RTTE (run-time-to-empty) of 65 minutes. This is designed to allow 25 minutes of run-time until "replace
battery" under nominal R Series operating conditions in monitor mode with no printer.

If the system shuts down due a slowly depleting battery, the R Series system requires user-intervention to turn back on. This could be due to battery drop-
in or the front panel switch.