R Series Service Manual

1-38

9650-0903-01 Rev. J

Warning!

NIBP transducer calibration can affect clinical readings of the NIBP parameter. Ensure that the NIBP Transducer Calibration
procedure is performed correctly, followed by an NIBP Monitor Test to verify proper operation.

22.6

On the R Series unit, press the Set High
softkey to calibrate the transducer to a
250 mmHg pressure reading.

The NIBP pressure transducer registers its voltage output at a known pressure
of 250 mmHg. The field adjacent to the 250 mmHg value changes to PASS.

Note: If the R Series displays a FAIL reading, verify the NIBP simulator’s

pressure setting and connection to the R Series and repeat the step.

o

o

22.7

On the NIBP simulator, set the pressure
parameter to stimulate a different cuff pressure
(for example, 205 mmHg).

The NIBP simulator displays the specified pressure reading.

22.8

On the R Series unit, press the Read Cuff
softkey.

Verify that the value displayed is accurate within
±3 mmHg of the pressure parameter set on the
NIBP simulator.

The NIBP module measures the pressure from the NIBP simulator and
displays the value in the Cuff Pressure field.

o

o

22.9

On the R Series unit, press the EXIT softkey
twice.

The R Series returns to the main NIBP Service Mode screen, then to normal
Monitor mode operation.

o

o

Record your results on the Maintenance Test Checklist.