Chapter 13 troubleshooting – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 495

Advertising
background image

9650-0912-01 Rev. K

ZOLL R Series Operator’s Guide

13–1

Chapter 13

Troubleshooting

The troubleshooting information provided in this chapter is intended for use by nontechnical
medical personnel during device operation. This chapter answers many of the common
problems or questions that may arise during operation.

If trouble persists after consulting this guide, contact the appropriate technical personnel or
ZOLL Technical Service Department. Refer to the R Series Service Manual for more detailed
troubleshooting information.

Code-Ready

Symptom

Recommended Action

1. Code Readiness indicator displays

a red “X” while unit is turned off.

Turn the unit on. Print Code Readiness log as described on
page 12-6. Follow messages and prompts to resolve the
problem.

2. Code Readiness indicator displays

a red “X” while unit is powered on.

Turn the unit off and then on again. Follow messages and
prompts to resolve the problem.

Perform a Manual Defibrillator test as described in “Manual
Defibrillator Testing” on page 12-3.

If the unit continues to fail, take it out of service and contact
ZOLL Technical Service.