Removing the NIBP Assembly
1. Disconnect the hose from the manifold and cut the
cable tie securing the pump to the chassis.
2. Disconnect the cable from the NIBP assembly from
parameter power supply by pushing down on latch at
the top of the connector.
Remove the 2 screws securing the NIBP assembly to
the chassis.
Note:
When reassembling, ensure NIBP tubing is not routed
between pump wiring. Secure the NIBP bracket to the
anchor with a cable tie.
Note:
During reassembly, torque to 6 in-lbs.