R Series Service Manual
1-18
9650-0903-01 Rev. J
9.0 Notch Filter Test
Tools Needed: Impulse 4000 (or equivalent).
Test Setup:
1. Connect the ECG cable to the Impulse 4000.
2.
Connect the ECG cable to the unit
.
Do this...
Observe this...
Pass/Fail/NA
9.1
Turn the selector switch to MONITOR mode (for
AED units turn to ON.)
9.2
Select lead I, size 3x.
Select 60Hz sine wave (or 50 Hz for a 50Hz
unit) on the Fluke 4000.
9.3
Press RECORDER button.
Verify that the waveform amplitude on the strip chart is less than 1.5 mm.
o
o
o
9.4
Turn the unit off.
Record your results on the Maintenance Test Checklist.