0 shock test, Shock test -25 – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 35

Advertising
background image

R Series Service Manual

9650-0903-01 Rev. J

1-

25

14.0 Shock Test

Tools Needed: Fluke Biomedical 4000 or equivalent defibrillator analyzer.

Test Setup:

1. Stop watch.

2.

Connect the OneStep cable via the adapter (D.N.I #3010-0378) to the defibrillator analyzer.

3.

Ensure that a fully charged battery is installed in the unit.

Do this...

Observe this...

Pass/Fail/NA

14.1

Turn the selector switch to DEFIB mode.(for
AED units, turn to ON, and select MANUAL
MODE).

14.2

Press the ENERGY SELECT down arrow until
5J displays.

DEFIB 5J SEL displays.

o

o

o

14.3

Press the CHARGE button. Wait for the
SHOCK button to illuminate.

DEFIB 5J RDY displays.

o

o

o

14.4

Press the SHOCK button.

Unit discharges 3J-7J into the simulator.

o

o

o

14.5

Press the ENERGY SELECT up arrow until 50J
displays.

Verify that DEFIB 50J SEL is displayed

o

o

o

14.6

Press the CHARGE button. Wait for the
SHOCK button to illuminate.

Verify that DEFIB 50J RDY is displayed.

o

o

o

14.7

Press the SHOCK button.

Verify that the unit discharges 46J-62J into the simulator.

o

o

o

14.8

Press the ENERGY SELECT up arrow until
100J displays.

Verify that DEFIB 100J SEL is displayed.

o

o

o

14.9

Press the CHARGE button. Wait for the
SHOCK button to illuminate.

Verify that DEFIB 100J RDY is displayed.

o

o

o

14.10

Press the SHOCK button.

Verify the unit discharges 93J-125J into the simulator.

o

o

o

Advertising