R Series Service Manual
9650-0903-01 Rev. J
2-13
SYSTEM FAULT 36
Filtered sum of all the supply voltages is out of range.
Replace analog board or digital board.
SYSTEM FAULT 37
1/2 scale reference voltage is out of range.
Replace analog board or PD engine.
SYSTEM FAULT 38
Failure to shutdown after "shutdown order" is written to the
RTC.
Replace digital board.
TEST FAILED
30J defib self-test failed.
Replace universal cable, paddles or high voltage
module, capacitor, and/or system board.
TRANSFERRING DATA
Transferring data message.
USE PADDLE DISCHG
Defibrillator is charged and front-panel discharge button
pressed when either external paddles or internal spoons
with discharge buttons are connected.
USE PADS TO PACE
MFE accessory other than Pads detected in Pace mode.
USER SETUP REQ.
Both copies of configuration data are bad, or software with
a configuration rev older than the current one was loaded.
Reconfigure unit.
VF ALARMS OFF
Alarms enabled in Pace Mode or when Continuous
Analysis active or started in manual mode but the current
lead is not Pads, Lead I, Lead II, or Lead III. Also displayed
if the Alarm button is pressed but the Heart Rate alarm is
disabled.
ZERO CO2 ADAPTER
Negative CO
2
detected. May be caused by a sensor that
was zeroed with CO
2
in the airway, or by an optical
blockage of the airway adapter.
Check the airway adapter and clean if necessary.
Perform a mainstream airway adapter zero as
described in “Zeroing the Mainstream
CAPNOSTAT 5 CO
2
Sensor/Airway Adapter.”
ZERO DONE
The sensor/adapter zero is finished.
No action required.
ZEROING CO2 ADAPTER
Adapter zeroing is in progress.
Wait for the adapter zeroing to finish.
Error Message
Explanation
User
Advisory
Technical Action