R Series Service Manual

1-10

9650-0903-01 Rev. J

Do this...

Observe this...

Pass/Fail

4.1

Turn the selector switch to MONITOR (for AED
units turn to ON.)

The unit should not turn on.

o

o

4.2

Turn the unit off.

4.3

Adjust the power supply voltage to 10.8V and
turn the selector switch to MONITOR.

The unit should turn on.

No LOW BATTERY message displays.

o

o

4.4

Low Battery Test

Set voltage to 10.5V.

LOW BATTERY message displays within 30 seconds.

o

o

4.5

Set voltage to 10.2V.

REPLACE BATTERY message displays within 30 seconds.

o

o

4.6

Turn the unit off.

Record your results on the Maintenance Test Checklist.

?

Battery Well

15 Amp
Supply

?

Red

Black