ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 40

Advertising
background image

R Series Service Manual

1-30

9650-0903-01 Rev. J

17.5

Increase the output to 80mA

Output on the Impulse 4000 is 80mA or +/- 5mA.

o

o

17.6

Increase the output to 120mA.

Output on the Impulse 4000 is 120mA or +/- 6mA.

o

o

17.7

Increase the output to 140mA.

Output on the Impulse 4000 is 140mA or +/- 7mA. Pulse width is 40mS +/-
2mS.

o

o

17.8

Decrease the output to 60mA.

Decrease the rate to 30 ppm.

Pacer rate on Impulse 4000 is 29-31 ppm.

o

o

17.9

Increase the rate to 180ppm.

Pacer rate on Impulse 4000 is 177-183 ppm.

o

o

17.10

Connect the ECG cable to the R Series and
Impulse 4000. Select the ECG at 60 BPM on
the Impulse 4000.

ECG at 60 BPM is seen on the display and no stimulus markers.

o

o

17.11

Press the Async Pace softkey.

ECG at 60 BPM seen on the display with the pace stimulus markers displayed.
Async pace message displays.

o

o

17.12

Turn off Impulse 4000. Set Pacer Rate to
100ppm. Press the RECORDER ON button.

Observe the pace stimulus markers every 15mm +/-1mm.

o

o

17.13

Press and hold 4:1 button.

Observe the pace stimulus markers every 60 mm+/- 1.5 mm.

o

o

Record your results on the Maintenance Test Checklist.

Do this...

Observe this...

Pass/Fail