Chapter 13 maintenance – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 307
9650-0904-01 Rev. K
ZOLL R Series Operator’s Guide
13–1
Chapter 13
Maintenance
Defibrillator equipment must be maintained to be ready for immediate use. The defibrillator
should be tested daily. The R Series defibrillator gives you two readiness testing options:
•
Automatic
•
Manual
Note:
Both automatic and manual test results are automatically recorded to internal memory.
When the R Series device ships from ZOLL, the Code Readiness indicator may show a red
“X.” A manual readiness test should be performed (in addition to other site-specific tests such
as HiPot and Leakage) prior to putting the device into service. Follow the procedure on page
12-3, and verify that the Code Readiness indicator displays a green “.”
Maintenance Frequency
Visual Inspection
Once per day, inspect Code Readiness indicator.
Code Readiness Test
(Automatic)
Once per day, unless configured Off, in which case, perform manual
defibrillator testing daily.
Manual Defibrillator
Testing
Once per week; daily if Code Readiness Test is configured Off.