R series service manual – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 161

Advertising
background image

R Series Service Manual

9650-0903-01 Rev. J

A-3

Delivered Energy at Every Defibrillator

Setting into a Range of Loads

* For all energy levels, accuracy is equal to either ±15% or 3 joules, whichever is greater.

Selected

Energy

Load

25

50

75

100

125

150

175

Accuracy*

1 J

1 J

1 J

1 J

1 J

1 J

1 J

1 J

±15%

2 J

1 J

2 J

2 J

2 J

2 J

2 J

2 J

±15%

3 J

2 J

3 J

3 J

3 J

3 J

3 J

3 J

±15%

4 J

3J

4 J

4 J

5 J

5 J

5 J

4 J

±15%

5 J

3 J

5 J

6 J

6 J

6 J

6 J

6 J

±15%

6 J

4 J

6 J

7 J

7 J

7 J

7 J

7 J

±15%

7 J

5 J

7 J

8 J

8 J

8 J

8 J

8 J

±15%

8 J

5 J

8 J

9 J

9 J

10 J

9 J

9 J

±15%

9 J

6 J

9 J

10 J

11 J

11 J

11 J

10 J

±15%

10 J

7 J

10 J

12 J

12 J

12 J

12 J

12 J

±15%

15 J

10 J

16 J

17 J

18 J

18 J

18 J

17 J

±15%

20 J

14 J

21 J

23 J

24 J

24 J

24 J

23 J

±15%

30 J

21 J

32 J

35 J

36 J

37 J

36 J

35 J

±15%

50 J

35 J

54 J

59 J

61 J

62 J

61 J

59 J

±15%

70 J

49 J

76 J

83 J

85 J

87 J

86 J

83 J

±15%

75 J

53 J

81 J

89 J

91 J

93 J

92 J

89 J

±15%

85 J

60 J

92 J

101 J

104 J

106 J

104 J

101 J

±15%

100 J

71 J

109 J

119 J

122 J

125 J

123 J

119 J

±15%

120 J

85 J

131 J

143 J

147 J

150 J

147 J

143 J

±15%

150 J

107 J

164 J

180 J

183 J

188 J

184 J

179 J

±15%

200 J

142 J

230 J

249 J

253 J

269 J

261 J

260 J

±15%