Repeat Analysis

After performing CPR for the configured period of CPR, the unit will restart ECG analysis
automatically.

Note:

Reanalysis of the ECG rhythm is inhibited for 3 seconds after a shock.

Continue Patient Care

Continue patient care according to medical protocols.

Operating Messages

The unit uses both audio and visual prompts to present critical information to operators. The
following information describes the unit default configuration. If your device has been custom
configured, some of the information may be different.

There are 10 voice prompts used in AED mode. Most of these prompts are accompanied by a
message displayed on the monitor. The voice prompts are given only once, but the monitor
continues to display the message until new action is taken by the operator or the device status
changes.

The unit will alternately display two different messages in the same field of the display when
two conditions are detected at the same time. For example, a LOW BATTERY message may
alternately display on the same line of the monitor as the CHECK PADS message.

Audio and Display Messages

The display messages and voice prompts that can occur during AED operation are described
below.

ATTACH PADS

If the unit is powered on without therapy pads connected to the patient, the ATTACH PADS
message is announced and displayed.

ANALYZING ECG/STAND CLEAR

The ANALYZING ECG message is displayed and the STAND CLEAR message is displayed and
announced when the ECG analysis starts automatically or after pressing the ANALYZE
softkey. They indicate that an active ECG analysis is in progress.

CHARGING XXXJ

ECG analysis is still in progress and a potentially shockable rhythm has been detected. The
current charge level is displayed.

SHOCK ADVISED

A shockable rhythm has been detected and defibrillation is advised. The selected energy level
has not yet been reached. The current energy level is displayed.