AED Operation

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

3–7

Once the unit is charged to the selected energy, the SHOCK button illuminates and the PRESS
SHOCK
message is announced and displayed.

Note:

Rhythm analysis does not continue after the defibrillator is charged and ready once a
decision to shock has been made. The R Series unit will not automatically disarm the
defibrillator if the patient’s rhythm reverts to a non-shockable rhythm before the shock
has been delivered.

A continuous tone sounds for 20 seconds, followed by an intermittent tone for 10 seconds.You
must deliver the shock within this 30 second interval, or the defibrillator will disarm itself.

Press and hold the illuminated SHOCK button on the front panel until energy is delivered to
the patient.

Observe the patient or ECG response to verify that the shock has been delivered.

After the energy has been delivered to the patient, the display returns to XXX J SEL.
SHOCKS:1
, indicating the number of shocks delivered to the patient.

Perform CPR

Begin chest compressions and rescue breathing per local protocol as prompted by the unit.

Note:

If OneStep CPR pads or OneStep Complete pads are connected, the unit monitors the
rate and depth of chest compressions and can display PUSH HARDER and GOOD
COMPRESSIONS
messages and voice prompts.