ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 60
R Series Service Manual
2-8
9650-0903-01 Rev. J
NIBP FAULT 263
Fault 90 received from NIBP module.
Replace NIBP module.
NIBP FAULT 264
Fault 91 received from NIBP module.
Replace NIBP module.
NIBP FAULT 265
Fault 97 received from NIBP module.
Replace NIBP module.
NIBP FAULT 266
Fault 98 received from NIBP module.
Replace NIBP module.
NIBP FAULT 267
Fault 99 received from NIBP module.
Replace NIBP module.
NIBP FAULT 268
Device no response after power up.
Replace NIBP module.
NIBP MEAS ABORTED
•
Cuff inflation pressure is set too high for attached cuff.
•
Inflation is too fast.
•
R Series is unable to find systolic value for 180
seconds.
•
Defibrillator is charged or charging.
•
User initiated abort.
•
Verify that you are using proper size cuff.
•
Check for cuff and hose blockage.\
•
Confirm that the unit was not charging.
•
If the problem persists, contact ZOLL
Technical Support.
NIBP NOT READY
•
The defibrillator is charged or charging in progress.
•
NIBP module is performing power-up self-test.
•
Wait until the unit discharges before taking the
next measurement.
•
Wait for more than 10 seconds after power-up
before taking blood pressure measurements.
NIBP OUT OF RANGE
The data from the NIBP module is out of range.
•
Check cuff fit and positioning. Switch cuff to
other arm.
•
Measure patient’s blood pressure with other
equipment.
•
If the problem persists, contact ZOLL
Technical Support.
NIBP SIGNAL WEAK
There is a weak or no oscillometric signal.
•
Check cuff fit and positioning.
•
Check for kinked hose.
•
Increase cuff inflation pressure if clinically
appropriate.
NO DATA TO TRANSFER
Summary Report / Device Check / Activity Log data
transfer error.
Error Message
Explanation
User
Advisory
Technical Action