ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 329
Defibrillator Specifications
9650-0904-01 Rev. K
ZOLL R Series Operator’s Guide
A–3
Messages
ERASING REPORT, INSERT CARD, REPLACE BATTERY,
LOW BATTERY, PERFORM CPR, NOISY ECG, RETRY ANALYSIS,
CHECK PATIENT, ANALYSIS HALTED, PRESS ANALYZE,
NO SHOCK ADV., SHOCK ADVISED, PRESS CHARGE, SELECT
PADS, SELECT ECG LEADS, SELECT DEFIB MODE,
VF ALARMS OFF, REMOVE SYNC, CHECK PADS, ATTACH PADS,
POOR PAD CONTACT, DEFIB PAD SHORT, PADDLE FAULT,
ECG LEAD OFF, USE PADDLE DISCHG, CANNOT CHARGE,
RELEASE SHOCK, PRESS SHOCK, 30J TEST OK, TEST FAILED,
PACER DISABLED, DEFIB DISABLED, SET PACE MA,
CHECK RECORDER, ANAYLZING ECG, FULLY RELEASE
Electrodes
Hands-free therapy
electrodes
•
OneStep electrodes
•
Pro
-
padz
•
Stat
-
padz
•
Pediatric pedi-padz
Defibrillator
Waveform
Rectilinear Biphasic
Energy selection (Adult)
Selection through front panel buttons or paddle buttons:
•
ADULT: 1, 2, 3, 4, 5, 6, 7 8, 9, 10, 15, 20, 30, 50, 75, 100, 120,
150, 200 joules
•
PEDIATRIC: 1, 2, 3, 4, 5, 6, 7 8, 9, 10, 15, 20, 30, 50, 70, 85, 100,
120, 150, 200 joules
Charge Time
•
Less than 7 seconds with a new, fully charged battery (first
15 charges to 200 joules).
•
For the fifteenth discharge at maximum energy, the charge time is
less than 10 seconds.
Depleted batteries result in a longer defibrillator charge time.
•
Less than 15 seconds when operating without a battery, using AC
power alone at 90% of the rated mains voltage.
•
Less than 25 seconds from the initial power on, with a new, fully
charged battery pack (depleted by up to fifteen 200 Joule
discharges) or when operating without a battery, using AC power
alone at 90% of the rated mains voltage.
•
Less than 30 seconds from initiation of rhythm analysis (advisory
mode) with a new, fully charged battery pack, (depleted by up to
fifteen 200 Joule discharges) or when operating without a battery,
using AC power alone at 90% of the rated mains voltage.
Patient Impedance
Range
15 - 300 ohms
Energy display
Screen shows selected and delivered energy.
Synchronized mode
Synchronizes defibrillator discharge to the patient’s R wave. SYNC is
indicated on the display with R wave markers above the ECG
waveform on the screen and stripchart. When ECG is monitored by
the device, meets the DF-80:2003 requirement of 60ms maximum
time delay between the peak of the R wave and the delivery of energy.
Charge controls
CHARGE button on front panel and apex paddle.
Paddles
Standard apex/sternum paddles. Adult plate slides off to expose
smaller plate for pediatric patients.