1. Getting Started

6

1.5 Getting Ready for Measurement

1.5.1 Connecting a Probe

Probe is calibrated together with the provided connection cable.

Please make sure the connection cable is connected when

performing a measurement.

* To change the length of a connection cable, recalibration

will be required.

1.5.2 Checking the Probe Number

Confirm that the number indicated on the one end of the probe

cable (Teflon cable) is consistent with the one indicated on the

screen of the instrument.

<<CAUTION>>

The probe number needs to be checked when multiple probes are purchased or the same probe is to be

used with multiple Anemomasters or a spare probe is used.

>>> Calibration data of the probe is written on characteristic ROM installed on the bottom of the

instrument. Please check the probe number because each probe has different characteristic.

1.5.3 To Attach and Detach the Probe Board

(2) Loosen up the screw.

>>> The screw will not come out.

(3) Pull it up vertically.

As the board is connected to the connector,
do not twist it or pull only one side.

>>> It may cause the connection failure.

(1) Open the bottom lid.

aaaaaaaaaa

Connection Cable
(heat-resistant: 80 °C)

Probe Cable
heat-resistant: 200°C

Probe Number

08.07.15

10:26:40

PROBE

46-H001

* This screen will be displayed when the

instrument is turned ON without the
probe connected to the instrument.

Probe Number
Are they consistent with each other?

Inside the bottom lid