3. Measurement Mode

17

START/STOP


(5)

ready

will be displayed indicating that the instrument is ready

to start a measurement.
Press START/STOP key to start sampling.

To go back to the measurement setting screen (a), press △

key.

[Start

Measuring]

[Waiting]

(6)

(6-1):

Sampling at one point.
(6-2):
After a sampling at one point is finished,
the instrument will standby for the next
sampling. On the screen, the average
value for one point will be displayed until
the next measurement will start.
(6-3):
One minute is left before the next
measurement starts
One minute later, the screen will
become screen (6-1) and a measurement
will start.

[Stop

Measuring]

(7) Repeat the above procedure from (6-1) to (6-3) for the

configured number of points.

To stop measuring halfway through: START/STOP or ▽ key.

- The data collected until the measurement is halted will be

calculated.

To stop measuring (1): △ key.

- It will go back to the screen (b) where the instrument is

ready to start a measurement.

To cancel a measurement (2): MENU key.

- It will go back to the menu screen.

Calculation

Result

Printing

Example

"

P24

Data at each point is an average data. Average, maximum and

minimum of the average data at each point will be displayed.

PRINT: hardcopy of the display screen

START/STOP or ▽: ready for measuring (screen b)

MENU: menu will be displayed

INT (min):

Set how often a measurement to be started.
Configurable sampling interval (min):

1 ~ 6, 10, 12, 15, 20, 30, 40, 50, 60, 100, 120, 150, 180 min.

* Interval (INT) has to be set longer time than sampling time at each point.

POINTS:

Set how many points you want to perform a measurement?
Configurable number of points: 1 ~ 999 points

* For S-TIME(S), DATA(N), MEMORY, PRINT, refer to page 15 describing the average measurement.

[

Measurement Condition]

(6-1)

<INT>

ready

T: 00/10 M:036
N: 000/010 P:016

16.8 m/s

127 ℃

<INT>

T: 01/10 M:036
N: 001/010 P:016

16.8 m/s

128 ℃

<INT>

WAIT 001 min
POINTS 001/024

16.8 m/s

126 ℃

<INT RESULT>

AVE: 19.8 m/s
MAX: 21.6 m/s
MIN : 18.5 m/s
AVE: 126 ℃
MAX: 129 ℃
MIN : 123 ℃

As for “ready” display,
refer to page 16
describing average
measurement.

(6-2)

(6-3)

(b)

WAIT 060 min

:

waiting time

remaining time length
before the next
measurement starts (min)

POINTS 001/024

: point

Current sampling
trial No.

Total set No. of

measurement trials

<INT>

WAIT 060 min
POINTS 001/024

12.8 m/s

126 ℃