Data output, 1 analog output – KANOMAX 6162 Anemomaster User Manual
Page 36
5. Data Output
30
5. Data Output
5.1 Analog Output
Output on the measurement screen -------------------------------------- ★ To be output every 0.25 second
★
Output Voltage: DC0 ~ 1V
★
Output Impedance: 47Ω
5.1.1 How to Change Air Velocity Range
Display Key
Procedure
MENU
(1) Press MENU key.
△
, ▽
SET
(2) Select 5. UTILITY.
Then press SET key.
△
, ▽
SET
(3) Select 2. ANALOG OUTPUT.
Then press SET key.
△
, ▽
SET
(4) Select 1. VELOCITY.
Then press SET key.
Output Range
Air Velocity
Air Temp.
0 ~ 5 m/s
0 ~ 50 °C
0 ~ 25 m/s
0 ~ 100 °C
0 ~ 50 m/s
0 ~ 200 °C
0 ~ 500 °C
Initial Screen:
Except when the screen is on hold
Measurement Mode: Except when the calculation result
is displayed or the instrument is
waiting for the next sampling
Analog Output Terminal
V:
Air Velocity
T:
Air Temperature
(a)
< MENU >
1.MONITOR
2.MEASUREMENT
3.DATA OUTPUT
4.MEMORY CLEAR
5
.UTILITY
< UTILITY >
1.CALENDAR
2
.ANALOG OUTPUT
3.PROGRAM SET
4.RS-232C
< ANALOG OUTPUT >
1
. VELOCITY
2. TEMPERATURE