KANOMAX 6162 Anemomaster User Manual

Page 41

Advertising
background image

5. Data Output

35

(1) Measurement Mode (VI: Air Velocity & Air Temperature)

(2)

Calculation Mode

(AVE: Average Measurement INT: Interval Measurement FLW: Flow Measurement)

(3) S-TIME (S) (Sampling Time)

(4) DATA (N) (Number of Data to be collected)

(5) POINTS (No. of samplings (partitions))

(6) Interval Measurement: INT(measurement interval)

Flow Measurement: AREA (m2) (Duct’s cross section area)

Output Format of Measurement Condition ----- M Command

Output Format of Measurement Data ----- T Command

<Command Explanation>

Command Function

Explanation

P

To check the number of

memory pages

After the command is received, AP will be returned.

Then, the number of pages stored will be output.

N

*

To check the number of

memory data

After the command is received, AN will be returned.

Then, the number of data in the specified page (*) will be output.

M

*

To check measurement

condition

After the command is received, AM will be returned.

Then, the measurement condition in the specified page will be output.

T

*

To output measurement

data

After the command is received, AT will be returned.

Then, the measurement data in the specified page will be ouput.

*

When page number is entered incorrectly, the error message E will be returned.

As for E1 and E3, refer to 5.2.3 To Transfer Raw Data (measurement data per second).

* AVG Mode:

dummy output

(1)

(2)

(3)

(4)

(5)

(6)

# # ; # # # ; # # ; # # # ; # # # ; # # # # # CRLF

: 25 Characters

VT ; FLW ; 60 ; 100 ; 10 ; 1. 456CRLF

(1)

(2)

(3)

(4)

(5)

(6)

[Output

Example]

23.6; 126; *** CRLF

air velocity, air temp, dummy

# # # # # ; # # # # # ; # # # # # CRLF :

19 Characters

air velocity air temp dummy

(Unit: m/s, ˚C, **)

[Output
Example]

Advertising