7

1.5.4 Display Screen – Monitor Screen

When the display does not change to the monitoring screen

Symptom

Solution

PROBE display remains.

Probe is not connected.

Ö

Turn the power OFF. Then after connecting the

probe, turn the power ON again.

The responsiveness of

reading is bad.


Isn’t SLOW1 or SLOW2 (displaying moving
average deviations) displayed on the upper right
corner of the screen?

Ö

Press FAST/SLOW key to switch to FAST.

Date and time are not

displayed.

Calculation program is set.

Ö

Press MENU key to select 1.MONITOR.

08.07.15 FAST

10:26:40

12.8 m/s

27.5 ℃

Power ON

This will be displayed for 2 sec.

Date ····· Year. Month. Date

Time(24-hour display) ····· Hour:Min:Sec

Š

Start measuring when this

screen is displayed.

When probe is not connected ) P6

Use FAST/SLOW key ) P9

* 3 options: FAST、SLOW、SLOW2

FAST: instantaneous value every second
SLOW1: moving average deviations for 5 seconds.
SLOW2: moving average deviations for 10 seconds.

* To change the date and time ) P12

Simultaneous display

Air velocity

Air temp.

08.07.15 FAST

10:26:40

PROBE

40-0001

VER. 1.0

KANOMAX

ANEMOMASTER

< AVE >

ready

T: 00/10 M:015
N: 000/010 P:007

12.8 m/s

27.5 ℃

08.07.15

10:26:40

PROBE

40-0001

08.07.15 SLOW1

10:26:40

12.8 m/s

27.5 ℃