3. Measurement Mode

15

, ▽

SET

(7) After measurement condition is set, select 5. SET OK!.

Then press SET key.

START/STOP

(8)

ready

will be displayed indicating that the instrument is ready

to start a measurement.

Press START/STOP key to start sampling.

To go back to the previous screen (measurement condition

setting screen) press △ key.

[Start

Measuring]

[Stop

Measuring]

[Calculation

Result]




Printing

Example

"

P24

To stop measuring halfway through: START/STOP or ▽ key.

- The data collected until the measurement is halted will be

calculated.

To cancel measuring (a): Press △ key.

- It will go back to screen (a) where the instrument is ready to

start a measurement.

To cancel a measurement (b): Press MENU key.

- It will go back to the menu screen.

PRINT: hardcopy of the display screen

START/STOP or ▽: ready for measuring (a)

MENU: menu will be displayed.

S-TIME (S): Set the length of sampling (instantaneous value) time.

Configurable sampling time (sec): 1 ~ 6, 10, 12, 15, 20, 30, 40, 50, 60 sec

DATA (N):

Set how many data to be taken every configured sampling time.
Configurable number of data: 1 ~ 6, 10, 12, 15, 20, 30, 40, 50, 60, 100, 120, 150, 180

MEMORY: When

YES, data will be stored in the internal memory.

For details on memory capacity, refer to P22.

PRINT:

When YES, the calculation result will be output to the printer automatically.
When NO, you can still print out the hardcopy of the calculation result by pressing PRINT key
after the calculation result is displayed.
Refer to P24 for printout samples.

[

Measurement Condition]

(a)

T:01/10···S-TIME(S)


M:015--- used memory (shown in %. When it shows 100 (%),

there is no memory left.)

N:001/060···DATA (N)

P:007--- the page number where the data is stored "P22

If 3.MEMORY NO is selected, it will not be displayed.

setting value (sec)

will be increased every second

setting value

will be increased every time data is collected.

3. MEMORY NO
4. PRINT NO

5

. SET OK!

<AVE>

ready

T: 00/10 M:015
N: 000/010 P:007

16.8 m/s

127 ℃

<AVE>

T: 01/10 M:015
N: 001/060 P:007

16.8 m/s

127 ℃

<AVE>

T: 10/10 M:016
N: 060/060 P:007

12.8 m/s

126 ℃

<AVE RESULT>

AVE: 12.8 m/s
MAX: 15.8 m/s
MIN : 10.8 m/s
AVE: 126 ℃
MAX: 129 ℃
MIN : 123 ℃