2. Basic Operation

9

2.2 How to Make the Reading More Readable

Š This function is useful when you want to obtain averaged readings as readings were volatile.

* This function is not available in Average Value Measurement Mode, Intermittent Operation Measurement Mode

and Air Flow Measurement Mode.

Mode

How to Take in Measurement Data

Explanation

FAST

Data is taken 4 times for one second. The

average value of the 4 data is displayed as

an instantaneous value every second.

SLOW 1

The average value for 5 seconds is

displayed every second.

Data shifts by 1 second.

SLOW 2

The average value for 10 seconds is

displayed every second.

Data shifts by 1 second.

0 5 10 15 20sec

(measuring time)

average for 1 sec

average for 5 sec

0 5 10 15 20sec

(measuring time)

average for 10 sec

0 5 10 15 20sec

(measuring time)

Every time you press FAST/SLOW key, the display at the

upper right corner changes in sequence.

Press FAST/SLOW key.

08.07.01 FAST

10:26:40

15.7 m/s

24.6 ℃

FAST:

Instantaneous value every second

SLOW1: Moving average deviation for 5 sec
SLOW2: Moving average deviation for 10 sec

FAST SLOW 1 SLOW 2