KANOMAX 6162 Anemomaster User Manual

Page 37

Advertising
background image

5. Data Output

31





, ▽

SET

(5)

(a) The range highlighted is the current set range.

If there is no change to be made, press SET key.

The screen will go back to the ANALOG OUTPUT screen (a).

(b) To change the range, select the range and press SET key.

(6) The selected range will be highlighted, and the screen will

automatically go back to the ANALOG OUTPUT screen (a).

MENU

(7) Press MENU key to go back to the menu screen.

* Analog voltage is output ony when the display screen indicates a

measurement is being performed.














To change the air temperature’s
output, perform the same
procedure.

(a)

< RANGE >

1. 0 5 m/s

2

.

0 25 m/s

3. 0 50 m/s

< MENU >

1

.MONITOR

2.MEASUREMENT
3.DATA OUTPUT
4.MEMORY CLEAR
5.UTILITY

< ANALOG OUTPUT >

1

. VELOCITY

2. TEMPERATURE

< RANGE >

1

.

0 5 m/s

2. 0 25 m/s

3. 0 50 m/s

< RANGE >

1.

0 5 m/s

2

. 0 25 m/s

3. 0 50 m/s

Range

Conversion

Equation

0~ 5 m/s

U = 0.005 × V

0~25 m/s

U = 0.025 × V

Air

Velocity

0~50 m/s

U = 0.05 × V

0~50 °C

T = 0.05 × V

0~100 °C
0~200 °C

T = 0.1 × V
T = 0.2 × V

Air

Temp.

0~500 °C

T = 0.05 × V

Velocity
Temp.

0
0

2.5

50.0

5.0 m/s
100°C

0.5

1.0

Out

p

ut

(V

)

<Example>

[Symbols]

U: Air

Velocity

(m/s)

T: Air

Temp

(°C)

V: Output

Voltage

(mV)

< Output Voltage Conversion>