KANOMAX 6162 Anemomaster User Manual
Page 19
2. Basic Operation
13
2.6 Printing Hard Copy of the Monitor Screen
2.6.1 What you need
★
Printer (sold separately) ····· Recommended model is DPU-201GS (Seiko Instruments Inc.)
★
Cable to connect the instrument and a printer (sold separately)
2.6.2 Printer Setting
Switch No.
Function
Anemomaster
Printer
SW 1
Word Length
8 bit
ON
SW 2
Parity Yes/No
None
ON
SW 3
Parity Setting
None
ON
SW 4∼6
Baud Rate
4800
Table Below
Baud Rate
SW 4
SW5
SW6
4800 OFF ON OFF
2.6.3 Signal Cable Connection
Anemomaster Printer
Signal Pin No.
Signal Name
Signal Pin No.
Signal Name
3 (orange)
TXD
3
DATA
1 (brown)
GND
5
GND
4 (yellow)
CTS
8
BUSY
2.6.4 Operation Procedure
1. Connect the instrument (RS-232C output terminal) and a printer.
2. Turn on both of the instrument and the printer.
3. Confirm that the Anemomaster’s display is an initial screen.
4. Press HOLD key to hold the display screen.
5. When the display is on hold, press PRINT key.
To Halt Printing Temporarily and To Cancel Printing
Temporary Halt: To stop printing temporarily, press PRINT key while printing.
To recommence printing, press PRINT key again.
Cancel Printing: To stop printing temporarily, press PRINT key while printing.
When the printing is halted, press MENU key.
As the menu screen will be displayed, select 1. MONITOR.
Press PRINT key when this
screen (on hold) is displayed.
08.07.01
hold
10:26:40
5.38 m/s
25.7 °C