8 counter (optional) 25, Üû/o – West Control Solutions CI 45 User Manual

Page 25

Advertising
background image

Displays

With configuration for O

2

measurement (see above), the oxygen content is displayed as process value with the selected

unit (see above) on line 1. Max. 4 characters can be displayed.

With display range overflow, “

EEEE” is displayed .

Example: the ppm range is selected, but the value is a % value.
When exceeding the display span start, 0 is displayed.

+

Tip: the unit can be displayed on line 2.

6.8

Counter (optional)

Digital input di1 can be configured as a counter input (dependent on ordering code).

The function operating as a pulse counter is set as

Up counter

(

ConF / Cn.Fr / I.Fnc = 1, 2 )

Down counter

(

ConF / Cn.Fr / I.Fnc = 3, 4 )

Active edge is configurable

g

The counter state is updated continuously in the background with the sample & hold function activated
(

r p. ) .

a

The counter state is not stored permanently. It is reset to the counter start value (Cnt.S) after supply
voltage switch-on.

Up counter function

The counter starts with start value Cnt.S,
adjustable via

ConF / Cn.Fr / Cnt.S.

With every edge at input di1, the counter is incremented by 1.

A counter end value can be defined via

ConF / Cn.Fr / Cnt.E.

Pulses in excess are not counted.
With the counter end value switched off, incrementing is done up to
the max. coun ter value.

g

Please, note that the counter end value must be higher than the
counter start value (Cnt.E > Cnt.S).

Down counter

The counter starts at counter start value Cnt.S,
adjustable via

ConF / Cn.Fr / Cnt.S.

The counter is decremented by 1 with every edge at input di1.

The counter end value can be defined via

ConF / Cn.Fr /

Cnt.E.
Pulses in excess are not counted.
With the counter end value switched off, decrementing is done down
to 0.

Functions

UNIFLEX CI 45

Counter (optional)

25

Cnt.S

Cnt.E

cnt

Cnt.E = off

Cnt.S

Cnt.E

cnt

Cnt.E = off

20.95

üû/o