Digital input di1, Thermocouples (table 1), Resistive sensors (table 2) – West Control Solutions CI 45 User Manual

Page 55: Current and voltage measuring ranges (table 3), Designed as

Advertising
background image

Thermocouples (Table 1)

Cold-junction compensation

- Internal,

Additional error:
typ.:
max

.:

ß_ 0,5 K

ß -2,5 K

- external,

- constant value

0...100 °C

Further technical data as INP1

Resistive sensors (Table 2)

Connection technique:

3-wire

Measurement span

Further technical data as INP1.

Current and voltage measuring ranges
(Table 3)

Remaining technical data as for INP1 except:
- Voltage input ranges –10 / 0...+10V and -5...+5V not possible.
- Millivolt input ranges: break monitoring always active.

DIGITAL INPUT DI1

Designed as:

a) Contact input

Connection of a potential-free contact suitable for switching “dry”
circuits.

Switched voltage:

5 V

Current:

1 mA

b) Opto-coupler input (optional)

For active control signals.

Rated voltage:

24 V DC external

Logic “0”:

-3 V … 5 V

Logic “1”:

15 V… 30 V

Current requirement :

max. 6 mA

Technical Data

UNIFLEX CI 45

55

Type

Sensor current

Measurement range

Error

Typical resol.

Pt100***

ß 0,25mA

-200...100(150)°C

-328...212(302)°F

ß 1 K

0.05 K

Pt100

-200...850°C

-328...1562°F

ß 2 K

0.05 K

Pt1000

-200...850°C

-328...1562°F

ß 2 K

0.05 K

KTY 11-6*

-50...150°C

-58...302°F

ß 2 K

0.05 K

Special*

0...4500 [**

ß 0.1%

0.005%

Special*

0...450 [**

ß 0.1%

0.005%

Potentiom.

0...160 [**

ß 0.1%

0.005%

Potentiom.

0...450 [**

ß 0.1%

0.005%

Potentiom.

0...1600 [**

ß 0.1%

0.005%

Potentiom.

0...4500 [**

ß 0.1%

0.005%

* Default setting is the characteristic for KTY 11-6 (-50...150°C)
** Including lead resistance
*** up to 150°C at reduced lead resistance (max. 160 [)

Table 2: Resistive inputs

Measurement range

Input resistance

Error

Typical resol.(

Æ)

0...10 V

~ 110 k[

ß 0.1 %

0.3 mV

-10...10 V

~ 110 k[

ß 0.1 %

0.6 mV

-5...5 V

~ 110 k[

ß 0.1 %

0.3 mV

-2,5...115 mV*

? 1 M[

ß 0.1 %

4 úV

-25...1150 mV*

?1 M[

ß 0.1 %

40 úV

-25...90 mV*

?1 M[

ß 0.1 %

4 úV

-500...500 mV*

?1 M[

ß 0.1 %

40 úV

-200...200 mV*

?1 M[

ß 0.1 %

20 úV

0...20 mA

20 [

ß 0.1 %

0.8 úA

* For INP1: high-impedance, without break monitoring

Table 3: Current and voltage input

Advertising