5 How to Operate

5-4. Modes

RM-5000

- 34 -













5-4. Modes


Details on each mode are provided as follows. (* Operations are slightly different depending on the model.)

Mode Item

LCD

display

Details

Detection
Mode

- Gas

concentration
Gas name

Normal state

Gas Alarm
Test Mode

- Gas

concentration

Perform the alarm test.

Zero Adjustment (Span
Adjustment)

1-1 ZERO
(1-1 SPAN)

Perform the zero adjustment.
(In case of oxygen 0 - 25 %, perform the span adjustment.)

Setting Display

1-2.CONFIRM

Show the setting of the typical menu.
• First alarm setpoint (AL1)
• Second alarm setpoint (AL2)
• Alarm delay time
• Zero suppression value
• Zero follower ON/OFF
• Indicator

type

Peak Value Display

1-3 PEAK

Display the peak concentration value when a gas is detected.

Main Unit Version Display

1-4 RM VER

Show the program version of the main unit.

AMP Version Display

1-5 AMP VER

Show the program version of the amplifier unit.

RS-485 Address Display

1-6 ADDRESS

Show the address.

RS-485 Communication
Setting Display

1-7 485 PTRN

Show the setting status of the communication function.

Maintenance
Mode
(User)

Regular Maintenance Mode
Switching

1-8 M MODE

Switch to the regular maintenance mode.

Gas Introduction Display

2-0 GAS TEST

Perform the gas introduction test in the regular maintenance mode.

Zero Adjustment

2-1 ZERO

Perform the zero adjustment.

Span Adjustment

2-2 SPAN

Perform the span adjustment.

Last Calibrated Date

2-3 LAST CAL

Show the last calibrated date.

Heater Current Display

2-4 CUR CAL

Show the heater current.

Environmental Setting 1

2-5 SETTING1

Operation setting
SE 0 INHIBIT setting (INHIBIT)
SE 1 Alarm setpoint value setting (ALM P)
SE 2 Alarm delay time setting (ALM DLY)
SE 3 Fault test (F TEST)

Environmental Setting 2

2-6 SETTING2

Functions setting
SE 0 Address setting (ADDRESS)
SE 1 Date/Time setting (DAY TIME)
SE 2 Zero suppression value setting (SUPPRESS)
SE 3 Zero suppression type setting (SUP TYPE)
SE 4 Contact setting for alarm test (TEST RLY)
SE 5 External output setting for alarm test (TEST4-20)
SE 6 Energized/De-energized setting (RLY PTRN)
SE 7 Alarm type setting (ALM TYP)
SE 8 Alarm pattern setting (ALM PTRN)
SE 9 Alarm value limiter setting (AL LIMIT)
SE10 Fault alarm pattern setting (FLT PTRN)
SE11 Zero follower ON/OFF setting (ZERO F)
SE12 External output in maintenance mode setting (MNT OUT)
SE13 External output adjustment (MA 4-20)

Maintenance
Mode
(Regular
maintenance)

Environmental Setting 3

2-7 SETTING3

Adjustment and setting
SE 0 Amplifier initialization (AMP DEF)
SE 1 Heater current adjustment (HEAT ADJ)
SE 2 Load voltage adjustment (LOAD ADJ)
SE 3 Measured gas selection (GAS SEL)
SE 4 Peak hold setting (PEAKHOLD)
SE 5 First alarm LCD setting (ALM1 LCD)

CAUTION

• Do not turn off the indicator/alarm unit during the initial clear.

• If a new sensor is installed or the sensor is replaced after the indicator/alarm unit is started, the

sensor must be warmed up for a specified period which is determined depending on the type of
the sensor. After the warm-up is completed, perform a gas calibration. Read also the operating
manual of the gas detector head.

• During the warm-up, the alarm activation and output signals are unstable. Provide a prior

notification to the related sections so that they can prepare for false abnormalities.