Maintenance, 1. maintenance intervals and items – RKI Instruments RM-5000 Series Indicator/Alarm Unit User Manual

Page 52

Advertising
background image

7 Maintenance

7-1. Maintenance intervals and items

RM-5000

- 52 -





7

Maintenance





The indicator/alarm unit is an important instrument for the purpose of safety.
To maintain the performance of the indicator/alarm unit and improve the reliability of safety, perform a
regular maintenance.

NOTE

To use the multi-unit case, also refer to the separate operating manual.


7-1. Maintenance intervals and items


• Daily maintenance: Perform maintenance before beginning to work.

• Monthly maintenance: Perform maintenance on the alarm circuit (alarm test) once a month.
• Regular maintenance: Perform maintenance once or more for every six months to maintain the

performance as a safety unit.

Maintenance

item

Maintenance content

Daily

maintenance

Monthly

maintenance

Regular

maintenance

Power Supply
Check

Check that the power lamp lights up.

{

{

{

Concentration
Display Check

Check that the concentration display value is
zero (or 20.9% on the oxygen deficiency
meter).
When the reading is incorrect, perform the zero
adjustment after ensuring that no other gases
exist around the detector head.

{

{

{

Alarm Test

Inspect the alarm circuit by using the alarm test
function.

{

{