2. fault alarm activation, Response to gas alarm – RKI Instruments RM-5000 Series Indicator/Alarm Unit User Manual

Page 47

Advertising
background image

6 Operations and Functions

6-2. Fault alarm activation

- 47 -

RM-5000

<Response to Gas Alarm>

In case of responding to a leaked gas

When a gas alarm is triggered, take actions in accordance with your management rules of gas alarm.
Normally, take the following actions.
• Check the reading of the indicator/alarm unit.

NOTE

If a gas leak is momentary, the reading may already have dropped when you check it. In addition, when
the alarm is triggered by noise or other incidental conditions other than a gas, the reading may have
already dropped.


• Based on your management rules of gas alarm, no one should be allowed to access the monitored zone

to ensure safety.

• If the Gas Concentration Display continues to be shown, close the main valve of the gas, and then check

that the gas concentration reading is dropped.

• Assuming that gases remain, wear protective equipment to avoid dangers and go to the gas leak point,

and then check if gases remain using a portable gas detector etc.

• If you can determine that the point is free from dangers, take actions to fix the gas leak.

6-2. Fault alarm activation


A fault alarm is triggered when the indicator/alarm unit detects abnormalities <<Auto-Reset>>.
After a fault alarm is triggered, the FAULT lamp (yellow) blinks and an error message is displayed on the
LCD. Determine the causes and take appropriate actions.
After the indicator/alarm unit is successfully returned from the fault, it restarts with the process normally
performed right after it is turned on (initial clear).
If the indicator/alarm unit has problems and is repeatedly malfunctioning, contact RIKEN KEIKI immediately.









* Display example: E-1 DETECTOR

sensor abnormality



NOTE

For information on malfunctions (error messages), see "9. Troubleshooting".