Environmental setting 2 "2-6 – RKI Instruments RM-5000 Series Indicator/Alarm Unit User Manual

Page 59

Advertising
background image

7 Maintenance

7-2. Regular maintenance mode

- 59 -

RM-5000

<Environmental Setting 2 "2-6">

In the environmental setting 2, specify the settings of functions. (* It is recommended that setting changes should be recorded in a log.)
The environmental setting 2 includes setting menus which are usually not used. Be careful not to change these settings by mistake.
<<Environmental Setting 2>>

During maintenance
During inhibit

Lamp on
Lamp off
Lamp blinking

2-6. SETTING2

Press SET key.

Date/Time Setting

=> P62

SE 0. ADDRESS

Set the RS-485 communication
address.(Model with an option
mounted)
Press the ▲ or ▼ key to change
the value, and then press the
SET key to confirm the value.

SE 1. DAY TIME

Set the RS-485
communication Set the
date/time of the main unit.

SE 2. SUPPRESS

Set the suppression value.
Press the ▲ or ▼ key to
change the value, and then
press the SET key to confirm
the value.

SE 3. SUP TYPE

Set the suppression type.
Press the ▲ or ▼ key to select
either CUT/SLOPE, and then
press the SET key to confirm
the selection. When CUT (Cut
Off) is selected, values
exceeding the suppression
value are directly displayed.
When SLOPE (Slope) is
selected, values exceeding the
suppression value are gently
displayed.

SE 4. TEST RLY

Set the contact activation for
an alarm test.
Press the ▲ or ▼ key to select
either ON/OFF, and then press
the SET key to confirm the
selection. When ON is
selected, the contact can be
activated during an alarm test.

SE 5. TEST4-20

Set the external output for an
alarm test.