Storage, relocation and disposal, 3. disposal of products – RKI Instruments RM-5000 Series Indicator/Alarm Unit User Manual

Page 73

Advertising
background image

8 Storage, Relocation and Disposal 8-1. Procedures to store the indicator/alarm unit or leave it for a long time

- 73 -

RM-5000





8

Storage, Relocation and

Disposal




8-1. Procedures to store the indicator/alarm unit or

leave it for a long time


The indicator/alarm unit must be stored under the following environmental conditions.
• In a dark place under the normal temperature and humidity away from direct sunlight

• In a place where gases, solvents or vapors are not present

• In a place free from vibrations or shocks

8-2. Procedures to relocate the indicator/alarm unit or

use it again


When the indicator/alarm unit is relocated, select a new place in accordance with “4-2. Precautions for
installation points” and “4-4. How to install”.
For information on wiring, see “4-5. How to wire". The unpowered time for the detector head (sensor) or
indicator/alarm unit must be minimized when the device unit is relocated.








8-3. Disposal of products


When the indicator/alarm unit is disposed of, it must be treated properly as an industrial waste in
accordance with the local regulations.

CAUTION

When using a relocated or stopped/stored indicator/alarm unit again, never fail to perform a gas
calibration. For information on readjustment including gas calibration, please contact RIKEN KEIKI.

Advertising