7 Maintenance

7-3. Gas calibration method

- 69 -

RM-5000

7-3. Gas calibration method


Perform a gas calibration on the detector head (sensor) connected to the indicator/alarm unit in each mode
(zero adjustment mode and span adjustment mode) using the calibration gas.
• Zero adjustment gas (collected in a gas sampling bag)

• Span gas (collected in a gas sampling bag)

• Gas

sampling

bags







<Zero Adjustment "2-1">

This is used to perform the zero adjustment on the detector head (sensor).







NOTE

Before starting the zero adjustment, supply the zero adjustment gas to the detector head (sensor) and
wait until the reading is stabilized.


<<Zero Adjustment>>

WARNING

After the adjustment is completed, never fail to press the MODE key to return to the detection mode.

WARNING

When the zero adjustment is performed in the atmosphere, check the atmosphere around the
detector head (sensor) for freshness before beginning the adjustment. If other gases exist, the
adjustment cannot be performed properly, thus leading to dangers when the gas leaks.

2-1. ZERO

Press SET key.

Present Concentration Value Display

Press the SET key to perform the
zero adjustment.

During maintenance
During inhibit

Lamp on
Lamp off
Lamp blinking

Zero Adjustment Completed

Press the SET key to perform the
The menu returns to

2-1. ZERO automatically.

* When the zero adjustment fails

Return to 2-1.ZERO

Return to 2-1.ZERO