5 How to Operate

5-5. Detection mode

RM-5000

- 36 -

5-5. Detection mode

<Gas Name and Full Scale Display>

Display a gas name, full scale value, etc. that have been set in advance.
































NOTE

NC-5001W offers two reading ranges (low and high ranges).
If the displayed combustible gas concentration rises above the full scale of the low range, the display is
automatically switched to the high range.
On the other hand, if the gas concentration drops below the full scale of the low range, the display is
automatically switched back to the low range.
The low range is indicated by lighting of the WL display, and the high range by lighting of the WH display,
informing the present reading range (low or high range).

Example
Target gas

:

Isobutane

Reading range

:

0-2000 ppm

/

0 - 100%LEL

Status display

:

WL (low range)

/

WH (high range)


NOTE

At a low temperature, the response of the LCD display may get slow down.

Full scale value

Gas concentration digital

display

Gas name

Gas concentration bar

meter display

Power lamp

CAUTION

A reading under zero is suppressed with the 10% FS suppression.
A reading that is 10% FS or more under zero is displayed as "-0.0", which prevents an accurate gas
detection and needs the zero adjustment. For information on the suppression function, see "6-4.
Other functions".