3 diode array spectrometer, Diode array spectrometer – Metrohm NIRS Analyzer PRO User Manual

Page 11

Advertising
background image

▪▪▪▪▪▪▪

9

wavelength range from 1100-1650 nm. (The optical switch may be heard moving when instrument
scans are taken, if the ambient noise level is low enough.)

2.3

Diode Array Spectrometer

The diode-array spectrometer is composed of 512
discrete InGAs detector diodes. It has no moving
parts.

The wavelength accuracy is intrinsically stable, and
requires no adjustment. Temperature is maintained
internally to assure consistent and accurate
response.

NOTE: There are no user-serviceable parts in the
Diode Array Assembly. Do not attempt to open,
disassemble, or service this unit. Any attempt at
service or opening will immediately void the
warranty.

Power Supply

This takes the incoming AC Mains voltage and steps it down to 24 VDC, as used by the instrument.

Cooling Fan

The fan circulates air internally to maintain consistent, controlled temperature to the spectrometer
and all internal components. This assures repeatable optical response.

Sampling Point

This may be either a Window Reflectance system (as shown here), a Fiber Optic Probe, or Direct
Light. The internal instrument interface is the same for all styles. A fiber optic bundle carries the light
to the sample. A small bundle carries the returning energy from the sample to the Optical Switch.

AC Mains Block

This is explained in the section on electrical Connection. All connections should be made in
compliance with local codes, by a qualified electrician. Always use safety lockout procedures when
performing electrical wiring.

Ethernet RJ-45 Connector

The Ethernet cable is plugged in on the instrument motherboard.

Control PCB

Instrument functions are controlled from this PC board. There are no user-serviceable items or
procedures in connection with this board.