Metrohm NIRS Analyzer PRO User Manual

Page 28

Advertising
background image

26

▪▪▪▪▪▪▪

8.

Slide the gland nut and sealing ferrule

back from the threaded area.

9.

Prepare a “fish” line which will be used

to pull the new cable into the enclosure.

This line will be used to pull the new
cable back into the correct position.

10.

Gently pull the old cable out of the

threaded header, guiding the fish line as
needed.

Do not pull the line all the way out – the
far end should remain where it can be
used to pull the new line into the
instrument enclosure.

This view shows the fish line after
pulling. Remove tape and old cable.

11.

Note the position of the nut, seal and o-

ring on the existing cable. Remove each
one and prepare to install on the new
cable.