12.

Install the parts as shown on the new

cable. This cable is a “North American”
type of cable, with white, black and
green conductors.

13.

Secure the new cable to the fish line,

and prepare to gently pull the new cable
into the instrument.

14.

Pull the fish line inside the instrument,

while feeding the new cable gently
through the threaded opening, and up
through the corner of the instrument
cabinet.

15.

Run the new cable to the AC power

block. Push down on each tab and
install the correct wires as shown, all the
way into the aperture.

After installation, give each wire a
gentle tug to assure that it is seated
properly. If the cable “pulls out”, verify
that the cable end is not distorted, and
try again.