Metrohm NIRS Analyzer PRO User Manual

Page 65

Advertising
background image

▪▪▪▪▪▪▪

63

4.

4. When the test is finished,

the split-screen display will
appear as shown. (The test
normally takes about 5
minutes in this
configuration, on the
Reference Channel.)

Double-click in the lower
right quadrant to see the
tabulated results.

5.

This is a typical set of

Performance Test results for
the Reference Channel. The
tabulated results show
these items:

Scan:

10 is the default value.

EOC:

Errors on communication – this will be “0” nearly all the time.

P-P:

Peak to Peak noise

Minimum:

Magnitude of lowest-going noise peak

Wavelength: Location of negative peak

Maximum: Magnitude of highest-going noise peak

Bias:

Baseline offset of noise scans

RMS:

Root-mean-square value of P-P

Gain:

Not used on this instrument