13 diagnostics, 1 window reflection, 1 performance test – Metrohm NIRS Analyzer PRO User Manual

Page 45: Diagnostics, Window reflection, Performance test

Advertising
background image

▪▪▪▪▪▪▪

43

13

Diagnostics

Diagnostics are provided to assure ongoing performance of the Analyzer PRO instrument. Upon each
connection to the instrument, Self-Test is performed automatically. The instrument monitors itself,
and will alert the operator in the event of malfunction.

It is wise to institute ongoing testing to monitor instrument performance and operation. The
frequency for this testing depends upon application and usage, and should be set by the System
Manager and other personnel who are preparing the system for use. This team of people will have
the best information on frequency of use, and the need for ongoing testing.

The tests are quite similar for each instrument configuration; however there are small differences in
how the standards are presented. Therefore this manual gives specific directions for each
configuration. Find the configuration that will be used, and follow the directions accordingly.

Tests are described below.

13.1

Window Reflection

First, verify that the Data Collection Method is correctly set up.

1.

Please verify that the

Sampling System appears as
shown below, in the area
marked by the box.

There is no need to set “Cell”
or Detector” as these
selections default to the
proper settings.

2.

Click on OK if setting up the DCM, then select it in the “Select Data Collection Method”

window as shown in the previous section of this manual.

If using the View DCM command, simply click on “Close”.

13.1.1

Performance Test

It is wise to run Performance Test to verify that the instrument is fully warmed up and operating
correctly. Follow these steps:

Advertising