48
▪▪▪▪▪▪▪
6.
Vision shows the spectrum of the
certified 80% Reflectance
Standard spectrum, plotted from
1100-1650nm.
Chick “OK” to plot the spectrum
of the correction.
7.
Vision plots the correction
spectrum in dark green. This
correction spectrum will be
applied to the internal reference.
Click “OK” to apply the
correction.
8.
Vision applies the correction,
shown in purple.
Click “Yes” to save the ERC to
the Diagnostic Database.
9.
Vision asks to upload the new
ERC (Reference Standardization)
to the Analyzer PRO instrument.
Click “Yes” to proceed.