Metrohm NIRS Analyzer PRO User Manual

Page 53

Advertising
background image

▪▪▪▪▪▪▪

51

7.

This test runs in

about two
minutes. Split-
screen results
appear as
shown.

Double-click on
the lower right
quadrant to see
full tabulated
results.

8.

There are four

wavelength
peaks used,
which serve to
characterize the
full wavelength
scale of the
instrument.

N.I.S.T. uncertainty is applied for rare-earth-type wavelength standards, shown as a
tolerance of +/-1.0 nm.

The Pass result is shown at the bottom of this report for each wavelength peak tested.

9.

If a printed copy is needed, click on “Print Report”.

A copy is automatically saved to the Vision Diagnostic Database.
Click on “Close Report” when finished.

NOTE: We advise removing the standards from the sample window when not in use. The intense
light beam can warm the standards considerably, which will change their wavelength response.
Please place the standards on the window only for current testing, and remove them promptly when
finished.