Metrohm NIRS Analyzer PRO User Manual

Page 30

Advertising
background image

28

▪▪▪▪▪▪▪

16.

Be sure the black o-ring is installed into

the groove on the sealing ring as
shown.

17.

Thread the gland nut and hand-tighten.

18.

Using the same 22 mm wrench, tighten

the gland nut to compress the seal
properly. Do not crush the cable. When
tight, the cable should not “give” when
tugged slightly.

19.

Tie-wrap the new cable to the Ethernet

cable internally, to secure it from
moving under vibration. Cut off the
loose end.