Metrohm NIRS Analyzer PRO User Manual

Page 47

Advertising
background image

▪▪▪▪▪▪▪

45

5.

When the test is finished,

the split-screen display will
appear as shown. (The test
normally takes 9-10 minutes
in this configuration.)

Double-click in the lower
right quadrant to see the
tabulated results.

6.

The tabulated results show

these items:

Scan:

10 is the default value

EOC:

Errors on communication this will be “0” nearly all the time.

P-P:

Peak to Peak noise

Minimum:

Magnitude of lowest-going noise peak

Wavelength: Location of negative peak

Maximum: Magnitude of highest-going noise peak

Bias:

Baseline offset of noise scans

RMS:

Root-mean-square value of P-P

Gain:

Not used on this instrument