Metrohm NIRS Analyzer PRO User Manual
Page 47
Advertising
▪▪▪▪▪▪▪
45
5.
When the test is finished,
the split-screen display will
appear as shown. (The test
normally takes 9-10 minutes
in this configuration.)
Double-click in the lower
right quadrant to see the
tabulated results.
6.
The tabulated results show
these items:
•
Scan:
10 is the default value
•
EOC:
Errors on communication this will be “0” nearly all the time.
•
P-P:
Peak to Peak noise
•
Minimum:
Magnitude of lowest-going noise peak
•
Wavelength: Location of negative peak
•
Maximum: Magnitude of highest-going noise peak
•
Bias:
Baseline offset of noise scans
•
RMS:
Root-mean-square value of P-P
•
Gain:
Not used on this instrument