Metrohm NIRS Analyzer PRO User Manual
Page 26
24
▪▪▪▪▪▪▪
mains cable is replaced, the same type of cross-section of cable must be used to assure a good seal.
The cables are carefully routed to prevent interference with operation of the Analyzer PRO.
When replacing the temporary AC mains cable, follow this procedure:
1.
Remove power from the instrument.
Both poles must be broken.
Use safety lock-out procedures if
applicable.
2.
Unplug or disconnect existing,
temporary AC mains cable.
3.
Remove the 32 screws holding the cover
in place on the instrument. Use a 7 mm
socket or nut driver.
NOTE: If using a powered removal tool,
set the torque on a low setting, to avoid
stressing the screws. Start at a low
torque, then increase enough to remove
the screws easily. Use a similar torque
when re-installing the screws. Avoid
cross-threading.