3 weld flange, Weld flange – Metrohm NIRS Analyzer PRO User Manual

Page 22

Advertising
background image

20

▪▪▪▪▪▪▪

This photo shows one installation on a process
pipe.

The Analyzer PRO instrument is supported by
metal straps from a support above the
instrument, using the mounting flanges. In this
case the blind plates were drilled to accept the
mounting bolt. Because the mounting flange
openings are sealed, there is no risk of
contaminants getting inside the Analyzer PRO
enclosure.

This photo was taken before the cables were
secured with tie wraps.

When the mouning is finished, secure the power cable and commmunications cable to prevent
damage. Be careful where they exit the Analyzer PRO, as a very tight bend here may cause cable
damage. It is wist to protect the cables from any passing traffic or possbile impact.

4.3

Weld Flange

The Weld Flange may be used to mount the instrument to ports where sample material is moving in a
continuous manner, and sampling will be consistent. Additional supports using the mounting plates
are required to hold the weight of the Analyzer PRO. While this flange may be welded, may
customers prefer to bolt it in place. Either method is acceptable, as long as sampling is not impeded.

The Weld flange uses the standard sanitary
sample interface and offers a large diameter
surface that can be welded onto the process.
The Analyzer PRO is then connected to this
flange.

Additional supports using the mounting plates
are required to hold the weight of the Analyzer
PRO.

Sample windows must be clean before final
installation.

Advertising