3 wavelength certification, Wavelength certification – Metrohm NIRS Analyzer PRO User Manual

Page 52

Advertising
background image

50

▪▪▪▪▪▪▪

13.1.3

Wavelength Certification

This test verifies that the wavelength scale of the Analyzer PRO meets parameters specified by the
U.S. National Institute of Standards and Technology (N.I.S.T.) for NIR wavelength response using rare-
earth standards.

1.

Click on “Analyzer PRO Setup/Diagnostic”, then “Run

Wavelength Certification”.

2.

The default number of scans is 10. The default

selection is “Wavelength Standard Analyzer PRO”.

Click “OK”.

3.

The Standards set contains a mini-CD

with the standards file on it. Click on this
file, then click “Open”.

Vision initiates the test.

4.

Vision begins taking an instrument reference spectrum.

5.

Vision requests that the user place the

wavelength standard onto the sample window.

Do not click “OK” yet.

6.

Verify that the serial number of the standard is

the same as that shown in the Vision prompt box.

Click “OK” when ready. Vision will start the test.

Advertising